30 EDXtreme User Instructions
4.1.3 Setup--About
The next soft key is the About. The About menu shows an assortment of information
about your Dillon instrument. This can be handy for maintaining calibration,
troubleshooting or determining if the firmware can be upgraded. Press this and access
the following soft key set (refer to Figure 4.1):
Device Press this to see software revision and dynamometer information.
Calib This soft key access the following soft key set:
Points- Press this key to display the calibration loading points. This is
useful information that can be keyed in manually in the event instrument
memory is ever accidentally overwritten or corrupted.
Re-Cal- Will show next re-calibration date and contact detail.
Print- Press this key to transmit all the calibration information available to
a serial printer or computer.
O.Load Lists the number of overloads that have occurred since the last calibration.
On Time- Actual hours when unit is on.
Zero Lists the current zero point compared to the calibration zero point. If the
zero point has moved significantly, this may indicate a serious overload
has occurred and the instrument should be returned for service.
Ntwrk Lists radio information; the enabled channel and identifier.
Dillon suggests that calibration data be recorded and saved.